Skip to main content
C.F.R.Title 42Chapter I—Public Health Service, Department of Health and Human ServicesSubchapter G—Occupational Safety and Health Research and Related ActivitiesPart 84—Approval of Respiratory Protective DevicesSubpart J—Supplied-Air Respirators

42 C.F.R. § 84.156 (2026)

Airflow resistance test; Type C supplied-air respirator, demand class; minimum requirements

Edition and currency
eCFR text as of Sept. 29, 2026. The text is the edition Syfert holds; for a past event, check the version in force on that date.
Other sources
CURRENTeCFR text as of Sept. 29, 2026

(a) Inhalation resistance shall not exceed 50 millimeters (2 inches) of water at an air flow of 115 liters (4 cubic feet) per minute.

(b) The exhalation resistance to a flow of air at a rate of 85 liters (3 cubic feet) per minute shall not exceed 25 millimeters (1 inch) of water.

Enactment and amendments

Syfert holds no separate history line for this section. Where this code prints a source or history note, it appears with the text, usually at the end.

Versions

Edition shown: eCFR text as of Sept. 29, 2026. Syfert holds no earlier or later version of this section.