C.F.R. homeSearch C.F.R.
C.F.R.Title 47Chapter I—Federal Communications CommissionSubchapter D—Safety and Special Radio ServicesPart 95—Personal Radio ServicesSubpart I—Medical Device Radio Communications Service

47 C.F.R. § 95.2563 (2026)

MedRadio frequency bands

Edition and currency
eCFR text as of Sept. 29, 2026. The text is the edition Syfert holds; for a past event, check the version in force on that date.
Other sources
CURRENTeCFR text as of Sept. 29, 2026

MedRadio transmitters operate in the 401-406 MHz, 413-419 MHz, 426-432 MHz, 438-444 MHz, 451-457 MHz, and 2360-2400 MHz bands. The FCC does not specify a channeling scheme for MedRadio systems.

(a) MedRadio transmitters associated with medical implant devices, which incorporate a frequency monitoring system as set forth in § 95.2559(a), may transmit on any frequency in the 401-406 MHz band.

(b) MedRadio transmitters associated with medical implant devices, which do not incorporate a frequency monitoring system as set forth in § 95.2559(a), may transmit on any frequency in the 401-402 MHz or 405-406 MHz bands, or on the frequency 403.65 MHz in the 402-405 MHz band.

(c) MedRadio transmitters associated with medical body-worn devices, regardless of whether a frequency monitoring system as set forth in § 95.2559(a) is employed, may transmit on any frequency in the 401-402 MHz or 405-406 MHz bands.

(d) MedRadio transmitters that are used externally to evaluate the efficacy of a more permanent medical implant device, regardless of whether a frequency monitoring system as set forth in § 95.2559(a) is employed, may operate on any frequency in the 402-405 MHz band, provided that:

(1) Such external body-worn operation is limited solely to evaluating with a patient the efficacy of a fully implanted permanent medical device that is intended to replace the temporary body-worn device;

(2) RF transmissions from the external device must cease following the patient evaluation period, which may not exceed 30 days, except where a health care practitioner determines that additional time is necessary due to unforeseen circumstances;

(3) The maximum output power of the temporary body-worn device must not exceed 200 nW EIRP; and

(4) The temporary body-worn device must comply fully with all other MedRadio rules applicable to medical implant device operation in the 402-405 MHz band.

(e) Only MedRadio transmitters that are part of a Medical Micropower Network (MMN) may operate in the 413-419 MHz, 426-432 MHz, 438-444 MHz, and 451-457 MHz bands. Each MedRadio transmitter that is part of an MMN must be capable of operating in each of the following bands: 413-419 MHz, 426-432 MHz, 438-444 MHz, and 451-457 MHz. All MedRadio transmitters that are part of a single MMN must operate in the same band.

(f) Only MedRadio transmitters that are part of a Medical Body Area Network (MBAN) may operate in the 2360-2400 MHz band.

Enactment and amendments

Syfert holds no separate history line for this section. Where this code prints a source or history note, it appears with the text, usually at the end.

Versions

Edition shown: eCFR text as of Sept. 29, 2026. Syfert holds no earlier or later version of this section.

C.F.R. Home  |  Search C.F.R.  |  Search cases for this section  |  syfert.com