If I am a user facility, what reporting requirements apply to me?

Edition and currency
eCFR text as of Sept. 29, 2026. The text is the edition Syfert holds; for a past event, check the version in force on that date.
Other sources
CURRENTeCFR text as of Sept. 29, 2026 cited 2001–2017 5citing cases
At a glance

Courts have cited this section in 5 opinions since 2001.

(a) You must submit reports to the manufacturer or to us, or both, as specified in paragraphs (a)(1) and (a)(2) of this section as follows:

(1) Reports of death. You must submit a report to us as soon as practicable but no more than 10 work days after the day that you become aware of information, from any source, that reasonably suggests that a device has or may have caused or contributed to the death of a patient of your facility. You must also submit the report to the device manufacturer, if known. You must submit the information required by § 803.32. Reports sent to the Agency must be submitted in accordance with the requirements of § 803.12(b).

(2) Reports of serious injury. You must submit a report to the manufacturer of the device no later than 10 work days after the day that you become aware of information, from any source, that reasonably suggests that a device has or may have caused or contributed to a serious injury to a patient of your facility. If the manufacturer is not known, you must submit the report to us. You must report information required by § 803.32. Reports sent to the Agency must be submitted in accordance with the requirements of § 803.12 (b).

(b) What information does FDA consider “reasonably known” to me? You must submit all information required in this subpart C that is reasonably known to you. This information includes information found in documents that you possess and any information that becomes available as a result of reasonable followup within your facility. You are not required to evaluate or investigate the event by obtaining or evaluating information that you do not reasonably know.

The 5 cases that cite 21 C.F.R. § 803.30 most heavily, weighted by court and treatment. RelevanceNewestTreatment
1
GREENBurger v. Lutheran Gen. Hosp., 759 N.E.2d 533 (Ill. 2001).
cites § 803.30 2×cited by 136 cases
Where it cites § 803.30
“21 (2001) (Medicare and Medicaid regulations require hospitals to engage in ongoing quality assurance reporting); 21 C.F.R. § 803.30 et seq. (2001) (failures of medical devices that result in death or serious injury to a hospital patient must be reported to the FDA); 77 Ill.”
The case’s holding
The right to privacy is not absolute and only unreasonable invasions of privacy are constitutionally forbidden.
Later courts cite it for
“the hospital is not a third party with respect to its own medical information, which is compiled by the hospital's own caregivers.” · 5 citing courts
2
GREENUniv. of Kentucky v. Bunnell, 532 S.W.3d 658 (Ky. Ct. App. 2017).
cites § 803.30 1×cited by 5 cases
Where it cites § 803.30
“” 21 C.F.R. § 803.30 (a) (1). ■ . Although the Guidance says “a provider should only place information in its PSES if it intends to report that information to the PSO[,]” Guidance at 32656, the administrators are pragmatists and recognize the possibility that "the provider is…”
3
GREENIn re Intuitive Surgical Sec. Litig., 65 F. Supp. 3d 821 (N.D. Cal. 2014).
cites § 803.30 1×cited by 5 cases
Where it cites § 803.30
“§ 360i(b)(l)(B); 21 C.F.R. §§ 803.30 , 803.50. If the hospital’s report reasonably suggests that the device may have contributed to a serious injury or death, or malfunctioned in such a way that it could have caused serious injury or death, then the manufacturer must also report…”
4
GREENIn re Mentor Corp. ObTape Transobturator Sling Prods. Liab. Litig., 632 F. Supp. 2d 1370 (M.D. Ga. 2009).
cites § 803.30 2×cited by 2 cases
Where it cites § 803.30
“, 21 C.F.R. § 803.30 . These reports are known as “user facility reports.”
5
NO FLAGBurger v. Lutheran Gen. Hosp., No. 89643, 89644 cons. Rel (Ill. Oct. 18, 2001).
cites § 803.30 1×cited by 0 cases
Where it cites § 803.30
“21 (2001) (Medicare and Medicaid regulations require hospitals to engage in ongoing quality assurance reporting); 21 C.F.R. §803.30 et seq . (2001) (failures of medical devices that result in death or serious injury to a hospital patient must be reported to the FDA); 77 Ill.”

The flag is Syfertize’s treatment of the citing case itself, not of the statute. “Where it cites” is the passage in the opinion; the holding is a summary of the whole case and may concern another point. Annotations are extracted automatically from the opinions in the Syfert caselaw corpus.

Citing cases by year
200120092017

One bar per year. Tallest: 2 cases in 2001. These are opinions in the Syfert collection, which holds more of some courts’ recent opinions than their older ones, so a rise can reflect what the collection holds and not more litigation.

By court
Ill.2N.D. Cal.1M.D. Ga.1Ky. Ct. App.1
By subsection

Courts have not cited this section by subsection, so its cases are not split up here. The Leading cases tab lists the cases that cite it most heavily.

Enactment and amendments

Syfert holds no separate history line for this section. Where this code prints a source or history note, it appears with the text, usually at the end.

Versions

Edition shown: eCFR text as of Sept. 29, 2026. Syfert holds no earlier or later version of this section.

C.F.R. Home  |  Search C.F.R.  |  Search cases for this section  |  syfert.com